2nd Floor Spectrophotometer w/ monitor
- Turn on power switch.
- Bring up Menu (hit Mode button?) at top of screen; it asks if you want to make changes? Y or N.
- Hit Yes.
- Hit the number 1. (for changing wavelength)
- Hit Enter.
- type in wavelength ex. 260
- Hit Enter. Now you're back at the main Menu which asks if you want to make changes ? Y or N.
- Hit Yes.
- Hit the number 5. (for changing "Processing" wavelength)
- Hit Enter.
- Hit the number 1 (for lambda 1)
- Hit Enter.
- type in wavelength ex. 260
- Hit Enter. Now you're back at the main Menu which asks if you want to make changes ? Y or N.
- Hit Yes.
- Hit the number 7. (for print data)
- Hit Enter.
- Place blank cuvette in rear slot.
- Place sample cuvette in the front slot.
- Spectrophotometer shows reading of sample at the lower right hand side of the screen.
- Hit Start/Stop to print out reading.
Instructions for the Gilford Stasar II Spectrophotometer
- Turn on Vacuum Pump. Switch is located under counter top.
- Purge air through system by holding down sample actuator bar for a few seconds.
- Aspirate blank at least twice and adjust zero. Check to see if reading is steady
- Aspirate sample at least twice and take reading.
- After reading last sample, purge air and aspirate Coulter Clenz (the blue crap) throughout the system (hold down sample actuator bar until you see blue liquid entering vacuum bottle)
- Turn off Vacuum Pump.